X-ray Tube Voltage & Current Meter AB-2015E
AB-2015E allows the simultaneous measurement of X-ray tube voltage and current for calibration and maintenance of X-ray equipment.
Excellent frequency response covers the full range of X-ray equipment, from inverter to single-phase type.
Excellent frequency response covers the full range of X-ray equipment, from inverter to single-phase type.
Features
■PC interface / Micro-Processor control / Auto zero-adjustment (One click of button)
■Self-checking function / User calibration (Digital calibration)
■Digital setting of sampling time (50/60Hz auto switch) and delay time (0~255ms)
■Display holding and exposure indicator / Connectable to Model AB-2150D-MⅡ*
* Needs modification and adjustment at manufacturer
■Self-checking function / User calibration (Digital calibration)
■Digital setting of sampling time (50/60Hz auto switch) and delay time (0~255ms)
■Display holding and exposure indicator / Connectable to Model AB-2150D-MⅡ*
* Needs modification and adjustment at manufacturer
Specifications
Measuring range
Tube voltage | Rated voltage: Less than 20kV - 150kV Max. kV on display: 188kV |
---|---|
Tube current-RAD | Rated current: 3 - 1500mA Max. current: More than 1900mA |
Tube current-FLU | Rated current: 0.03 - 15.00mA Max. mA on display: More than 19.00mA |
Measuring accuracy
Tube voltage | Less than 40kV: Less than ±0.7kV 40-100kV: Less than ±1.5%±0.1kV *Traceable up to 100kV |
---|---|
Tube current | Radiography: ±1%±1mA Fluoroscopy: ±3%±0.01mA |
Measuring time
Set range | Single phase: 2P, 1P (Auto switch for 50/60 Hz) Inverter: 1, 2, 4, 8, 16, 32ms |
---|---|
Set accuracy | More than 10ms: less than ±2% of set value Less than 10ms: less than ±200μs |
Monitor output
Anode voltage | 20kV/V |
---|---|
Cathode voltage | 20kV/V |
Tube voltage | 20kV/V |
Tube current | Radiography: 200mA/V Fluoroscopy: 2mA/V |
Data output
Format | Serial character data (ASCII Code) |
---|---|
Data type | Tube voltage and tube current (CSV) |
Communication type | Standard: RS-232C Protocol: No control sequence Delimiter: <CR> <LF> |
Communication details | Baud rate: 9600 Baud (fixed value) Data bit length: 8 bits Parity: N (none) Stop bit: 1 bit |
General specifications
Power source | Voltage: 100V ±10% Frequency: 50±1 or 60±1 Hz |
---|---|
Power consumption | Less than 50 VA (Non-measuring time) |
Environment | Ambient temperature: 5 - 35℃ Humidity: 45 -85% Ambient pressure: 70 -106kPa |